| Medical Device Problem Codes |
Fracture (1260); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Pain (1994); Nonunion/Delayed-union Bone Fracture (2369)
|
| Date of Event |
03/18/2015
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
One (1) unit - 901 185 - 800 010 691 lot.Two (2) units - 901 183 - lot 3608694.One (1) unit - 901 183 - lot 3618229.The screws and plate were produced in accordance with the specified / projected.There is no evidence that the failure has occurred by material fail.Conclusion: the material failure occurred by the use of only four (4) screws in the distal part of the plate (900.502).And also because of the distal femur of the patient compromised by the presence of the knee prosthesis and intramedullary bone cement and releasing the load before signs of consolidation display region.
|
| |
|
Event or Problem Description
|
|
In a surgery performed, a 5-hole left less invasive distal femur plate was implanted, approximately two months after the procedure the patient returns for another surgery because the distal femur plate was loose and had broken fixing screws.
|
| |
|