| Catalog Number |
CM3162N |
| Medical Device Problem Code |
Air Leak (1008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is currently being evaluated; smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.Device evaluation in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that the balloon of a portex® bivona® custom 6.5 a/c standard ns tracheostomy tube was leaking after one week in use.The leak was noticed by a private duty nurse during bedside care.Cuff patency was tested prior to use, and the cuff was filled with air.Velcro tracheostomy tube ties were used to hold the device in place.The nurse changed the tracheostomy tube.No injury was reported, and the event was reported as resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for evaluation.Functional testing involved leak testing and found no signs of leaking.A review of the device history record found that the lot had no abnormalities and was within specified dimensions.Functional testing involved dimensional testing and found the cuff passed the wall thickness specification.Based on the evidence, a root cause was unable to be confirmed.No fault was found with the device.
|
| |
|