• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II BIFURCATED STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETBF3620C145E
Medical Device Problem Codes Leak/Splash (1354); Malposition of Device (2616)
Health Effect - Clinical Code Injury (2348)
Date of Event 04/11/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 5.2 cm in diameter abdominal aortic aneurysm.The aortic neck was 10 mm in length and 27 mm in diameter at the renal arteries.There was a sharp curve/turn in the distal part of the aortic neck.It was reported that during the index procedure the physician was modelling the bifurcated stent graft with a reliant balloon and inadvertently pulled the endurant bifurcated stent graft distally.The inaccurate placement of the endurant bifurcated stent graft resulted in a proximal type i endoleak.The physician elected to implant an endurant aortic cuff and the events were resolved.Per the physician, the event was related to the patient¿s anatomy of angulation of the distal aortic neck.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
Corrected information: sex.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6542008
Report Number2953200-2017-00735
Device Sequence Number615947
Product Code MIH
UDI-Device Identifier00613994991058
UDI-Public00613994991058
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2018
Device Model NumberETBF3620C145E
Device Catalogue NumberETBF3620C145E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/11/2017
Supplement Date Received by Manufacturer04/11/2017
Initial Report FDA Received Date05/03/2017
Supplement Report FDA Received Date10/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-