• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH FITMORE HIP STEM, UNCEMENTED, A/5, TAPER 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT)22

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH FITMORE HIP STEM, UNCEMENTED, A/5, TAPER 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT)22 Back to Search Results
Model Number N/A
Device Problem Device Slipped (1584)
Patient Problems Pain (1994); Scar Tissue (2060)
Event Date 01/09/2014
Event Type  Injury  
Manufacturer Narrative
The manufacturer did not receive devices, x-rays, or other source documents for review as the patient has not been revised.Where lot numbers were received for the devices, the device history records were reviewed and found to be conforming.A cause for this specific event cannot be ascertained from the information provided.Should additional information become available and an investigation result be available, that changes this assessment, an amended medical device report will be submitted.This is a split complaint with zimmer inc.(b)(6) reported on (b)(4).Zimmer¿s reference number of this file is (b)(4).
 
Event Description
It was reported that the patient was implanted a fitmore, hip stem, uncemented, a/5, taper 12/14 on (b)(6) 2013.The patient underwent a procedure to remove scar tissue and adhesions approximately four months post implantation after experiencing pain, slippage and instability.No devices were removed.
 
Manufacturer Narrative
As soon as additional information become available and/or an investigation result be available, an amended medical device report will be submitted.Zimmer's reference number of this file is (b)(4).
 
Event Description
It was further reported that the associated biolox head was revised due to pain and instability on (b)(6) 2014.The fitmore, hip stem, uncemented, a/5, taper 12/14 stayed implanted.A further event for the same patient leading to a revision surgery of this stem is reported in case (b)(4).
 
Manufacturer Narrative
No trend considering the following event is identified: instability.Device history records (dhr): the device manufacturing quality records indicate that the released components met all requirements to perform as intended.Event summary: it was reported that patient underwent total hip procedure on (b)(6) 2013 and experienced pain, slippage and instability due to scar tissue and adhesions.The first surgery was (b)(6) 2014 and removed scar tissue, no devices were removed.No medical data such as x-rays, surgical notes or any other case-relevant documents received.No product was returned to zimmer biomet for in-depth analysis as the device was not explanted.The compatibility check was performed from (b)(6) and showed that the product combination was approved by zimmer biomet.Root cause determination using dfmea(fitmore stem): malfunctioning of the device due to wrong handling of device due to unclear information.Possible: it is unknown if the surgeon followed the given surgical technique.Conclusion summary: it was reported that patient underwent total hip procedure on (b)(6) 2013 and experienced pain, slippage and instability due to scar tissue and adhesions.The first surgery was (b)(6) 2014 and removed scar tissue, no devices were removed.As no devices were removed, no devices were available for the investigation.Therefore the condition of the components is unknown.Neither x-rays, operative notes nor office visit notes were received.Based on the given information and the results of the investigation, the complaint could not be confirmed as the alleged failure could not be identified or reproduced.The need for corrective measures is not indicated and zimmer (b)(4) considers this case as closed.Zimmer¿s reference number of this file is cmp-(b)(4).
 
Event Description
It was yet reported, that during the surgery of (b)(6) 2014, scar tissue was removed only.The biolox® delta, ceramic femoral head, m, 2/0, taper 12/14 was not revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FITMORE HIP STEM, UNCEMENTED, A/5, TAPER 12/14
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL UNCEMENTED ACETABULAR COMPONENT)22
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6543319
MDR Text Key74356880
Report Number0009613350-2017-00613
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup,Followup
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/31/2021
Device Model NumberN/A
Device Catalogue Number01.00551.105
Device Lot Number2605633
Other Device ID Number00889024482753
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/12/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/05/2011
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight89
-
-