Both catheters (sn: (b)(4)) were returned to ekos for evaluation on 4/17/17.Ekos qa found that drug lumens on both iddcs (part of the catheter) contained blood and appeared to be occluded with blood.For the catheter with sn: (b)(4), the iddc (part of the catheter) was kinked and some fibrous material was present, and all drug holes in the iddc being occluded by blood and fibrous material.Ekos qa was unable to infuse fluids through the drug holes.For the catheter with sn: (b)(4), the evaluation found all drug holes in the iddc (part of the catheter) were also occluded by blood.Ekos qa was unable to infuse fluids through the drug holes.Delay in connecting the catheters to the infusion pump after the catheter replacement could have likely contributed to the instances.Delayed connection allows blood to back up in the iddc lumen and occlude drug infusion holes.However ekos was not able to determine the root cause of the instance as the customer reported.Device history record of both catheters was reviewed, and it shows that both catheters were manufactured per standard processes and met all acceptance criteria.
|
Two 135cm / 12cm ekosonic catheters (sn: (b)(4)) were used to treat a patient with pulmonary embolism on (b)(6) 2017.Downstream occlusion alarm (s) occurred at drug port when the catheters were placed into the patient.Trouble shooting was performed and did not resolve the alarm, then the treating physician replaced the ekosonic catheters with standard infusion catheter(s).The treatment was completed, and the patient was doing well.This mdr is to respond to the customer filed mdr (mdr number: (b)(4)).
|