The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of reay1529 showed no other similar product complaint(s) from this lot number.
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint that the catheter was sheared off during insertion was confirmed and was determined to be related to use of the device.The proximal segment of one 18g powerglide catheter, which consisted of the purple hub and green strain relief sleeve, was returned for investigation.The catheter segment extended 4.6cm from the distal end of the green sleeve.The distal segment of the catheter was not returned for investigation.A microscopic examination revealed that the catheter was scored within the lumen near the distal tip, which is typical of damage caused by a needle.A crescent shaped slit was observed at the distal end of the catheter segment.The adjoining surfaces of the crescent shaped slit were noted to be glossy and striated.The cross section at the distal tip of the catheter was dull and slightly granular.The break in the catheter appeared to be a result of complications with the insertion process.The damage to the catheter appeared to have been initiated with the sharp edge of the needle tip.The product ifu warns, ¿once the catheter has been advanced, do not re-insert the needle back into the catheter or pull the catheter back onto the needle.This may result in damage to the catheter.If the catheter needs to be repositioned, either do so without the aid of the needle, or remove both the catheter and the needle as a unit to prevent the needle from damaging or shearing the catheter.¿ the ifu also states, ¿do not perforate, tear, or fracture the catheter with the needle or guidewire during the procedure.¿ no further action is required because the complainant event could not be related to a deficiency in product manufacture or deficiency while under correct product use.
|