(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It should be noted mini trek rx, global, instruction for use states: do not use, or attempt to straighten, a catheter if the shaft has become bent or kinked; this may result in the shaft breaking.Instead, prepare a new catheter.The investigation determined the reported difficulties appear to be related to crcumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
|