• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT MINI TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT MINI TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012269-08
Device Problems Detachment Of Device Component (1104); Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that when the mini trek 1.50 x 8 mm was ready to cross the lesion the device bent.An attempt was made to put it back in the original way but it broke.The guiding catheter was removed where the broken piece was and there were no adverse patient effects.A new mini trek 1.5 x 8 mm was used to successfully complete the procedure.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It should be noted mini trek rx, global, instruction for use states: do not use, or attempt to straighten, a catheter if the shaft has become bent or kinked; this may result in the shaft breaking.Instead, prepare a new catheter.The investigation determined the reported difficulties appear to be related to crcumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6545583
MDR Text Key74514545
Report Number2024168-2017-03925
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue Number1012269-08
Device Lot Number50918GA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
-
-