Catalog Number 0006439
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 04/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Immucor technical support used a remote electronic connection method on (b)(6) 2017 to assess the instrument test well image in question, which was visually negative.An immucor field service engineer (fse) visited the customer site to assess the testing instrument in question on 20apr2017.The fse found the instrument to be operating as expected.The immucor laboratory tested retention product on 01may2017, which performed as expected.The customer site's full and complete telephone extension (b)(6).
 
Event or Problem Description
On (b)(6) 2017, a (b)(6) customer site reported an unexpected negative antibody screen when using capture-r ready-screen (i and ii) on a galileo neo instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-SCREEN (I AND II)
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key6545655
Report Number1034569-2017-00088
Device Sequence Number13705564
Product Code KSZ
UDI-Device Identifier10888234000631
UDI-Public10888234000631
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Medical Technologist
Type of Report Initial
Report Date (Section B) 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Expiration Date05/16/2017
Device Catalogue Number0006439
Device Lot NumberX463
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2017
Initial Report FDA Received Date05/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age34 YR
-
-