MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INGENIO; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INGENIO; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number V173
Device Problem Failure to Capture (1081)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/09/2017
Event Type  Injury  
Manufacturer Narrative
Analysis of the returned product is currently ongoing.This report will be updated upon completion of the analysis.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and left ventricular (lv) lead were explanted due to high pacing thresholds and intermittent capture at max output.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGENIO
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6546219
MDR Text Key74471580
Report Number2124215-2017-07864
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/21/2014
Device Model NumberV173
Other Device ID NumberINVIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received05/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/26/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4136; 4471; 4549; V173
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age76 YR
-
-