A system diagnostic test was performed and showed high impedance.It was also reported that patient had been manipulating the device by rotating the generator in the surgical pocket.X-ray images were reviewed by the physician and reportedly confirmed a lead fracture had occurred.The patient had a full revision surgery to replace the fractured lead and generator, and the post-op impedance values were within normal limits.While in surgery, it was reported that the lead was extremely twisted and coiled up.The explanted lead and generator have not been received by the manufacturer to date.
|
Analysis was completed on the returned lead portions, which did not include the electrodes.During the visual analysis, the outer silicone tubing had a twist and compressed appearance.Additionally, several abraded openings were identified on the lead tubing and the quadfilar coil was found to be broken approximately 3 mm from the electrode bifurcation.Scanning electron microscopy (sem) was performed on the broken coil and was unable to identify the coil fracture type.The abraded openings on the outer silicone tubing provided a leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.The set screw marks found on the lead connector pin provided evidence that at one point in time, there was a good mechanical and electrical connection between the lead and generator.Other than the abraded insulation and the lead fractures identified in the lab, the condition of the returned lead was consistent with those typically seen following explant.
|