| Catalog Number |
33820 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
04/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
System performance indicators were acceptable at and around the time of this event.However, the available information reasonably suggests that the system has malfunctioned as information from competitor devices suggest the beckman coulter device did not perform as intended.There is no indication that the device was returned for evaluation.
|
| |
|
Event or Problem Description
|
|
The customer reported obtaining reproducibly elevated tsh (thyroid stimulating hormone) results for one patient on the laboratory's unicel dxi800 access immunoassay system (serial number (b)(4)).The results were determined to be discordant with reproducibly lower tsh results obtained for the same patient by alternate methods (roche, siemens and abbot).Customer-provided data indicate that elevated access tsh resutls were also generated for this patient on an alternate unicel dxi800 access immunoassay system (serial number (b)(4)) and access2 immunoassay system (serial number (b)(4)).The customer reported that the patient had been inappropriately diagnosed with and treated for hypothyroidism due to the elevated erroneous tsh results.The customer indicated that the patient had been inappropriately prescribed the maximum dosage of levothyroxine.The customer did not indicate sample type but did relate that samples were centrifuged for 10 minutes at 3500 rpm (rotations per minute).The customer reported that tsh qc (quality control) results generated on the system had been acceptable.
|
| |
|
Search Alerts/Recalls
|