MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1040-060
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Visual and functional inspections were performed on the returned device.The inflation issue and irregular balloon fold was not confirmed.It may be possible that the kinked inner member contributed to the inflation issue while in the anatomy preventing the balloon from inflating; however, this could not be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the common iliac artery.A 4.0x60mm armada 35 balloon catheter was advanced to the lesion.An attempt to inflate the balloon was made; however, the balloon dilated irregularly and the positioning part was not dilated (the dilated part was about 10mm distal to the positioning part).The device was removed and it was confirmed that the balloon was fold at the front end portion against the balloon marker.The procedure was successfully completed without additional treatment.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6549481
Report Number2024168-2017-03980
Device Sequence Number16396410
Product Code LIT
UDI-Device Identifier08717648154522
UDI-Public(01)08717648154522(17)171031(10)40521G1
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2017
Device Catalogue NumberB1040-060
Device Lot Number40521G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2017
Initial Report FDA Received Date05/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-