| Model Number |
60-03-75 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump console.The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue.The touch screen was replaced but it did not resolve the reported malfunction.The service representative replaced the entire centrifugal pump console to resolve the reported issue.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.The investigation is on-going.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova (b)(4) received a report that the display of the centrifugal pump console was blank during setup.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|