AV-TEMECULA-CT RX ACCULINK CAROTID STENT SYSTEM Back to Search Results
Catalog Number 1011337-30
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Date of event has been estimated.The device was not returned for analysis.It should be noted that the rx acculink instruction for use states: the rx acculink carotid stent system, used in conjunction with the abbott vascular embolic protection system specified below, is indicated for the treatment of patients at high and standard risk for adverse events from carotid endarterectomy who require carotid revascularization.The investigation was unable to determine a cause for the reported difficulty.The lot history record (lhr) for this product could not be reviewed and a similar incident query could not be performed because the product was not returned for evaluation and the lot number was not reported.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the popliteal artery.There was an issue deploying the acculink carotid stent system, once it reached the target lesion.The device was used off-label in the incorrect anatomy.A 5 x 4 mm acculink was used to complete the procedure.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX ACCULINK CAROTID STENT SYSTEM
Common Device Name
CAROTID STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6552977
Report Number2024168-2017-04006
Device Sequence Number628456
Product Code NIM
UDI-Device Identifier08717648076190
UDI-Public(01)08717648076190
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P040012
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1011337-30
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/20/2017
Initial Report FDA Received Date05/08/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-