PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARD® FLUORO-4¿ SILICONE URETERAL STENT Back to Search Results
Catalog Number 233626
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "warning: this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: for single use only.Do not re-sterilize.Do not use if package or product is damaged.Improper handling technique can seriously weaken the stent.Acute bending or overstressing during placement could result in subsequent separation of the stent at the point of stress after a prolonged indwelling period.Exercise care.Tearing of the stent can be caused by sharp instruments.Choice of stent size and duration of indwelling time are at the discretion of the physician.All stents may be subject to varying degrees of encrustation when placed in the urinary tract.Periodic checks of the stent by cystoscopic and/or radiographic means are recommended.When, at any time during the indwelling duration, encrustation is of sufficient severity that there is potential for occlusion of the stent or the patient experiences pain or discomfort which the physician determines to be associated with the presence of the stent, or if there is indication of infection in the area of the stent, the stent should be removed and, if the patient¿s condition permits, replaced with a new stent.Care should be exercised when removing the stent so as not to cause tearing or fragmentation.With any ureteral stent, migration is a possible complication which could require medical intervention for removal.Selection of too short a stent may result in migration.Multi-length ureteral stents: formation of knots in multi-length ureteral stents may occur.This may result in injury to the ureter during removal and/or the need for additional surgical intervention.The presence of a knot should be considered if significant resistance is encountered during attempts at removal.Directions for use: (open end).Prior to removing guidewire from hoop, inject sterile water through port to activate lubricious coating.Pass the guidewire flexible tip beyond the obstruction to the renal pelvis.Note: tortuosity in the obstructed ureter can often be resolved using the guidewire and open end ureteral catheter in combination.Pass the stent over the guidewire through the cystoscope, advancing it into the ureter with the push catheter under direct vision.Assistant holds the guidewire in position to prevent advancement of the wire into the renal parenchyma.Watch for the distal end of the stent at the ureterovesical junction.Stop advancement of the stent at that point.Assistant removes the guidewire as the operator holds the stent in position with the push catheter.The retention coil will form spontaneously.Carefully remove the push catheter from the cystoscope.Note: final adjustment, if necessary, can be made with endoscopic forceps.Stents can be removed easily by gentle withdrawal traction using endoscopic forceps.Fluoroscopy facilitates stent placement; however, standard radiography may be used.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable local, state and federal laws and regulations." (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that there was a split in the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® FLUORO-4¿ SILICONE URETERAL STENT
Common Device Name
URETERAL STENT
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
MDR Report Key6556848
Report Number1018233-2017-02320
Device Sequence Number700464
Product Code FAD
UDI-Device Identifier10801741011686
UDI-Public(01)10801741011686
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K823487
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number233626
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2017
Initial Report FDA Received Date05/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-