| Catalog Number |
6942-7-060 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales representative via phone that during a primary surgery for a fractured hip, that the taper head would not lock onto the stem.The physician tried 3 times.The physician used a bipolar head the same size and it worked.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding alleged "seating/locking issues" involving a unitrax c-taper sleeve was reported.The event was not confirmed.Method & results: device evaluation and results: the device was returned in used condition.Review of the device with a material analysis engineer indicated "damage consistent with in-service use was observed on the sleeve.In-service use can potentially affect the dimensions of the product." a dimensional inspection was performed on the returned device and found the device to be within specifications.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for this lot.Conclusions: the event could not be confirmed nor the root cause could be determined.Review of the device with a material analysis engineer indicated "damage consistent with in-service use was observed on the sleeve.In-service use can potentially affect the dimensions of the product." however, a dimensional inspection was performed on the returned device and found the device to be within specifications.
|
| |
|
Event or Problem Description
|
|
It was reported by the sales representative via phone that during a primary surgery for a fractured hip, that the taper head would not lock onto the stem.The physician tried 3 times.The physician used a bipolar head the same size and it worked.
|
| |
|