MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNITRAX C-TAPER SLEEVE -3MM; HIP JOINT FEMORAL (HEMI-HIP) METALLIC CEMENTED OR UNCEMENTED PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNITRAX C-TAPER SLEEVE -3MM; HIP JOINT FEMORAL (HEMI-HIP) METALLIC CEMENTED OR UNCEMENTED PROSTHESIS Back to Search Results
Catalog Number 6942-7-060
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Bone Fracture(s) (1870); No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported by the sales representative via phone that during a primary surgery for a fractured hip, that the taper head would not lock onto the stem.The physician tried 3 times.The physician used a bipolar head the same size and it worked.
 
Additional Manufacturer Narrative
An event regarding alleged "seating/locking issues" involving a unitrax c-taper sleeve was reported.The event was not confirmed.Method & results: device evaluation and results: the device was returned in used condition.Review of the device with a material analysis engineer indicated "damage consistent with in-service use was observed on the sleeve.In-service use can potentially affect the dimensions of the product." a dimensional inspection was performed on the returned device and found the device to be within specifications.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: review of the device history records indicates that all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other events for this lot.Conclusions: the event could not be confirmed nor the root cause could be determined.Review of the device with a material analysis engineer indicated "damage consistent with in-service use was observed on the sleeve.In-service use can potentially affect the dimensions of the product." however, a dimensional inspection was performed on the returned device and found the device to be within specifications.
 
Event or Problem Description
It was reported by the sales representative via phone that during a primary surgery for a fractured hip, that the taper head would not lock onto the stem.The physician tried 3 times.The physician used a bipolar head the same size and it worked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITRAX C-TAPER SLEEVE -3MM
Common Device Name
HIP JOINT FEMORAL (HEMI-HIP) METALLIC CEMENTED OR UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key6559001
Report Number0002249697-2017-01494
Device Sequence Number13486288
Product Code KWL
UDI-Device Identifier07613327013450
UDI-Public(01)07613327013450(11)170103(17)220131(10)58263802
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K992570
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Catalogue Number6942-7-060
Device Lot Number58263802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/12/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/10/2017
Supplement Report FDA Received Date06/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-