| Catalog Number |
03L82-21 |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot/serial number was manufactured prior to udi compliance date; therefore, only a di is provided evaluation of the customer issue included a review of the complaint text, a search for similar complaints, and labeling review.No adverse trend was identified for the customer issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics complaint investigation, no product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer observed discrepant falsely low glucose results generated using the clinical chemistry glucose reagents.The following data was provided and was generated in (b)(6) 2016.Initial 0.19 g/l (19 mg/dl); repeat 1.75 g/l (175 mg/dl); repeated as part of reproducibility, 10 replicates ranged from 1.26 to 1.87 g/l (126 to 187 mg/dl).No impact to patient management was reported.
|
| |
|
Search Alerts/Recalls
|