Catalog Number 72200616
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
A visual inspection was performed on the product and no issue was observed.A functional evaluation revealed that the motor overheated during testing.The complaint has been confirmed and a root cause has been associated with a mechanical component failure.Factors, unrelated to the manufacturing and design of the device which could have contributed to the reported event, include increased current draw due to corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time.Manufacturing records show that the device was released into distribution new in march of 2016.(b)(4).
 
Event Description
It was reported that the dyonics mdu hand control powermax was overheating.It was plugged into the shaver box before the case and the scrub felt it overheating.There were no patient injuries as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6559547
MDR Text Key75023398
Report Number1643264-2017-00262
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/05/2017
Initial Date FDA Received05/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse