Catalog Number CDF10020430
Medical Device Problem Codes Kinked (1339); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event was reported to codman cnv on (b)(6) 2017; however, did not meet mdr reporting criteria until 05/11/2017 when the analysis was completed.Conclusion: the deltapaq was returned for analysis.As seen through the returned packaging, the coil was found to be entangled around the dpu.The proximal end of the resistive heating coil section has been severely buckled (unreported).The proximal end of the coil had two compression and buckling sections of damage (unreported).No manufacturing defects were found.Due to post-procedural handling, cleaning, and packaging, it cannot be determined when and how all the unreported damages to the coil and dpu found occurred.A review of the manufacturing documentation associated with this lot presented no issues during the manufacturing or inspection process that can be related to the reported complaint.The complaint of the coils resistance in the headway microcatheter could not be confirmed.Without the return of the headway microcatheter and due to post-procedural handling, cleaning, packaging which may have produced further if not all damages found to the dpu and coil, the root cause of the coils resistance inside the microcatheter cannot be determined; however, the evidence suggests interference from an unknown source and location may have contributed to the event.Since there was no evidence to suggest the event was related to a manufacturing issue, no corrective actions will be taken at this time.This is an initial/final mdr report.This is 1 of 3 mdrs being submitted for this complaint, with associated report numbers of 2954740-2017-00105, 2954740-2017-00104 and 2954740-2017-00103.
 
Event or Problem Description
As reported by a healthcare professional, a physician felt strong resistance during advanced a deltaplush (cpl10015330/ c40127), a deltapaq (cdf10020430/ c33041) and a presidio (pc410073030/ c41431) through a headway 17 microcatheter.Upon return of the devices for analysis, the deltaplush and deltapaq coils were found to be compressed and buckled and the presidio coil was stretched.Other coils were able to be advanced.The physician replaced the coils with same like products and the procedure was successfully completed.There were no potential adverse events.After multiple attempts to obtain additional information, no additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAPAQ - CERECYTE MICROCOIL
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel
SZ 
MDR Report Key6559782
Report Number2954740-2017-00104
Device Sequence Number614225
Product Code HCG
UDI-Device Identifier00878528007081
UDI-Public(01)00878528007081(17)200131(10)C33041
Combination Product (Y/N)N
Initial Reporter CountryKN
PMA/510(K) Number
K080437
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Catalogue NumberCDF10020430
Device Lot NumberC33041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/11/2017
Initial Report FDA Received Date05/11/2017
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HEADWAY 17 MICROCATHETER
-
-