MAUDE Adverse Event Report: AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B1140-040
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/14/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: visual, dimensional and functional inspection was performed on the returned device.The reported resistance and separation was confirmed.It is likely that during use, the clearance between the guide wire and guide wire lumen of the armada 35 became reduced causing the reported resistance.Additional movement of the catheter against resistance likely resulted in the device become frozen together and during the attempt to remove the armada 35, the separation occurred.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, there is no indication of a lot specific product issue.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a highly stenosed lesion in the vena cava artery.A 14 x 40 mm armada 35 balloon separated while inside the patient.The separated portion was not recovered.Additionally, resistance with an unknown guide wire was felt when advancing and withdrawing the device.It is unknown as to how many atmospheres the balloon was inflated because it was inflated with hand injections.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6560307
Report Number2024168-2017-04064
Device Sequence Number10916696
Product Code LIT
UDI-Device Identifier08717648155024
UDI-Public(01)08717648155024(17)200430(10)61102G1
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 05/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberB1140-040
Device Lot Number61102G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2017
Initial Report FDA Received Date05/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient Weight125
-
-