| Model Number |
SL18P |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
02/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated.When the investigation is complete a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Staff noticed blood leak and hole was noted on venous side of catheter just below hub.
|
| |
|
Additional Manufacturer Narrative
|
|
The contract manufacturer reviewed the device history record, sample and leak testing process.It was noted that the lot of catheters in question was 100% leak tested and that all catheters passed prior to release.It was determined that the failure occurred after being released from their facility.
|
| |
|
Search Alerts/Recalls
|