• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH EXCEL; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH EXCEL; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B5LT
Device Problems Break (1069); Detachment Of Device Component (1104); Material Fragmentation (1261); Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/03/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Manufacturer Narrative
(b)(4).Batch # unk.The analysis results found that the b5lt instrument was received with the sleeve broken and melted.In addition, the tyvek was returned for analysis.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH EXCEL
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 00969
Manufacturer Contact
milton garrett
4545 creek road ml 120a
cincinnati, OH 45242
5133378865
MDR Report Key6561974
MDR Text Key75162547
Report Number3005075853-2017-02529
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2021
Device Catalogue NumberB5LT
Device Lot NumberP4R25U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2017
Was the Report Sent to FDA? No
Date Manufacturer Received05/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-