| Model Number |
EVOLUTPRO-23-US |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/19/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product analysis: the product remains implanted, therefore no product analysis can be performed.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the valve was deployed into a failed non-medtronic surgical aortic valve.The valve did not fully expand and a balloon aortic valvuloplasty (bav) was performed.Subsequently, the bav embolized the valve into the aorta.The valve was snared and stabilized.A second valve was implanted.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|