WELCH ALLYN KLEENSPEC 590 SERIES DISP. VAGINAL SPECULA Back to Search Results
Model Number 59001
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Laceration(s) (1946); Vaginal Mucosa Damage (2124)
Date of Event 03/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer did not provide a lot number therefore welch allyn was unable to review the device history record for the speculum.As the device and packaging was disposed of by the user we were unable to determine if this was a single rail or double rail speculum.Fourteen representative samples were returned to welch allyn for evaluation.The specula were a mix of single rail and double rail specula.Upon visual inspection of all 14 returned specula no damage was observed.Microscopic examination of the single rail specula showed no damage or cracking.The single rail specula were last manufactured on 30 july 2014.The enhanced double wall specula design was introduced in august 2014.Based on this information, no further action is required.
 
Event or Problem Description
Welch allyn received a report stating, kleenspec 590 series disposable vaginal speculum broke while being used on a patient during a pelvic exam.The speculum was examined by the physician prior to use and no visible cracks were noted.The physician inserted the speculum into the patient's vagina and immediately upon opening of the speculum, the upper blade shattered.Approximately half of the left side of the upper blade broke off of the speculum.The broken pieces were removed from the patient's vagina.The patient sustained an approximate 1 to 2 cm vaginal wall laceration from the sharp edge of the broken blade.Pressure was applied and the bleeding subsided.The patient recovered without further treatment or incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KLEENSPEC 590 SERIES DISP. VAGINAL SPECULA
Common Device Name
VAGINAL SPECULA
Manufacturer (Section D)
WELCH ALLYN
4341 state street road
p.o. box 220
skaneateles falls NY 13153 0220
MDR Report Key6562897
Report Number1316463-2017-00004
Device Sequence Number10989301
Product Code HIB
UDI-Device Identifier00732094143676
UDI-Public00732094143676
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K070964
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number59001
Device Catalogue Number59001
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2017
Initial Report FDA Received Date05/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-