| Model Number |
96530-017 |
| Medical Device Problem Codes |
Disconnection (1171); Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
01/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon receipt at medtronic's quality laboratory, visual inspection of the opened device showed the fitting was not connected to the cannula.After investigation at medtronic and vention, the complaint was confirmed, the connector was not bonded to the cannula body.This is the result of a manufacturing error.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during unpackaging, the customer observed the fitting was not connected to the body of the cannula.The product was replaced with another and not used for a case.There was no patient involvement in this product issue.
|
| |
|