The reporter indicated the surgeon implanted a 13.7mm micl13.7 implantable collamer lens, -8.5 diopter, in the patient's left eye (os), on (b)(6) 2017.The lens was explanted on (b)(6) 2017 due to excessive vaulting, closed angle and shallow anterior chamber.The pressure was 34mmhg but was brought down to normal with diamox.The lens was exchanged for a shorter lens.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch report will be submitted.
|