| Model Number |
48921002 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; while using the pilot cutter, the surgeon cut the channel and while back slapping the device out, the pressed fit back slap mechanism on the pilot cutter flew off onto the floor.Surgeon did not use a lot a force and it came off, leaving the pilot cutter stuck in the patient.Vice grips were applied back slap hammer was utilized and pilot cutter was removed.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection, device history review, complaint history review, risk assessment.The customer reported event was confirmed via the received image and visual inspection.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.The most likely cause of the reported event was determined to be related: (1) normal wear due to long time usage; (2) impaction overloading.
|
| |
|
Event or Problem Description
|
|
It was reported that; while using the pilot cutter, the surgeon cut the channel and while back slapping the device out, the pressed fit back slap mechanism on the pilot cutter flew off onto the floor.Surgeon did not use a lot a force and it came off, leaving the pilot cutter stuck in the patient.Vice grips were applied back slap hammer was utilized and pilot cutter was removed.There were no adverse consequences to the patient.
|
| |
|