ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER EDGEPLUS VALVED ENTRY SYSTEM; CANNULA, TROCAR, OPHTHALMIC Back to Search Results
Catalog Number 8065751585
Device Problem Leak/Splash (1354)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported leaky trocar valves during surgeries.Additional information and product sample have been requested for this case.No updates have been received at this time.
 
Manufacturer Narrative
No sample has been returned for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.An internal investigation has been opened to address reports of a similar nature.Due to an observed increased trend for this event, an internal investigation was initiated to determine if corrective and preventative actions were necessary.A root cause for the increased complaint rate was found to be related to a manufacturing post assembly inspection practice.(this complaint does not identify a reported lot and no sample was returned for evaluation so it cannot be determined if the complaint can be attributed to the post assembly inspection.) the post assembly inspection has been discontinued and an effectiveness check has been established and will be monitored periodically.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDGEPLUS VALVED ENTRY SYSTEM
Type of Device
CANNULA, TROCAR, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
nadia bailey
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key6567731
MDR Text Key75314725
Report Number2028159-2017-02004
Device Sequence Number1
Product Code NGY
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751585
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-