Catalog Number 7N8399
Device Problems Reflux within Device (1522); Failure to Infuse (2340)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that no flow was observed with a one-link needle-free iv connector.The reporter stated that when they attached the tubing and the pump read ¿occlusion¿.When the nurse attached the syringe and flushed again, there was some blood backed up in the tubing.The one-link set was removed and the solution flowed normally.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE-LINK NEUTRAL LUER ACTIVATED DEVICE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - AIBONITO
aibonito
Manufacturer (Section G)
BAXTER HEALTHCARE - AIBONITO
rd 721 km 0 3 po box 1389
aibonito 00705
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6568182
MDR Text Key75211344
Report Number1416980-2017-04209
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132734
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7N8399
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
EXTENSION SET; PUMP; SALINE; SYRINGE
Patient Age20 MO
-
-