| Model Number |
18785S |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received a report the label was missing description sizes and diameter measurements customer indicated there was no patient involvement with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The sample was received for analysis and the reported issue was confirmed.A visual inspection was performed and it was observed the printing on the pouches is missing information such as: part number, expiration date, dimensions of the tube and lot number (although a label from the outer box was received to identify it).Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report the label was missing description sizes and diameter measurements customer indicated there was no patient involvement with this event.
|
| |
|
Search Alerts/Recalls
|