FISHER & PAYKEL HEALTHCARE LTD COSYCOT INFANT WARMER; FMT Back to Search Results
Model Number IW934
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint iw934 infant warmer was inspected at the hospital by a trained fph service technician from our service centre in (b)(4).Results: during testing it was noted that the unit's power fail alarm did not function when the unit was disconnected from power.The root cause was identified to be the failure of a capacitor on the pcb board.Conclusion: the subject infant warmer was released for distribution in 2011.It is likely that the replaceable super capacitor on the pcb board has simply worn out as the unit is six years old.The super capacitor's function is to store sufficient electrical charge to power the warmer's visual and audible alarms in the event of a failure of mains power.Part of fisher & paykel healthcare's quality control process involves testing the power failure alarm of every warmer on the production line for functionality prior to distribution.The device technical/service manual contains a checklist which specifies that users perform safety, performance and functional checks including the power fail alarm at least once a year.Included in the warmer's maintenance checklist is a test of the super capacitor to ensure it is operating within specifications.The reported malfunction was discovered during a maintenance check with no patient involvement.The infant warmer was repaired and a new capacitor was fitted to the pcb.The subject infant warmer was returned to the customer after passing all the necessary safety and performance tests.Repaired at regional office.
 
Event Description
A hospital in (b)(6) requested a performance check of an iw934 cosycot infant warmer.During assessment it was found that the power fail audio alarm had failed.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COSYCOT INFANT WARMER
Type of Device
FMT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
jon stevens
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key6571134
MDR Text Key75329872
Report Number9611451-2017-00435
Device Sequence Number1
Product Code FMT
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K971695
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIW934
Device Catalogue NumberIW934
Device Lot Number111123
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/19/2017
Initial Date FDA Received05/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/23/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-