No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
It was reported that during deployment of the stent the health care provider allegedly experience resistance.It was further reported that the catheter was removed without incident and was identified to be broken.Reportedly, a pta balloon was then used to continue the procedure, and angioplasty was performed.There was no reported patient injury.
|
|
Manufacturing review: the device history records have been reviewed with special attention to the subassemblies, material review reports, raw material testing, manufacturing process, and quality control inspection.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Visual/microscopic inspection: the sample was returned.The safety clip was returned.The touhy borst valve was tight.The stent/graft had been partially deployed approximately 5.2cm.The outer catheter was broken approximately 0.8cm from the strain relief.The outer catheter was viewed under 10x magnification and was seen to be stretched on both sides of break and was received with proximal portion overlapping distal portion of catheter at break.Bent struts were noted under 10x magnification in the deployed portion of the stent graft and the outer layer of the graft was torn near the bent strut.No bent struts noted to un-deployed portion of graft.No other anomalies were noted to the stent graft.Functional/performance evaluation: guidewire patency was tested without issue.Further functional testing could not be performed, as the delivery system was returned with the outer catheter broken.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: a break in the outer catheter reinforcement area was noted and the stent graft was partial deployed from the system.Therefore, the investigation was confirmed for deployment failure due to the catheter break.The definitive root cause could not be determined based upon available information.It was unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) state: indication for use: flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Warnings: the stent graft (implant) cannot be repositioned after total or partial deployment.Once partially or fully deployed, the flair endovascular stent graft cannot be retracted or remounted onto the delivery system.Device removal after deployment can only be done with a surgical approach.Precautions: the safety and effectiveness of this device have not been established when used around a tight bend in a looped av graft.It is recommended to access the av graft at the venous side of the av graft or at the apex.Careful attention should be paid to ensure the device is appropriately sized to the actual graft diameter, taking into account any change to the stated graft diameter that may have resulted from previous interventions.The appropriate length device(s) should be selected so that the stent graft extends beyond the stenosis into at least 10 mm of the non-diseased graft towards the arterial inflow and into the non-diseased vein approximately 10 mm beyond the stenosis.During deployment of the stent graft, the entire length of the delivery system should be kept as straight as possible in order to mitigate the potential for distal stent graft misplacement.After full stent graft deployment, wait a few seconds to allow for complete device expansion before removing the delivery system over the guidewire.Stent graft dislodgement has been reported during removal of the delivery system; therefore, careful attention should be paid during this portion of the procedure to prevent such occurrences.In case of placement of two stent grafts (overlap), use the same device diameter in both cases.If a flared device is used to overlap, do not deploy the flared end inside the first stent graft.Ensure a minimum 10 mm overlap of the devices.Potential complications and adverse events: stent graft specific events that could be associated with clinical complications include stent graft misplacement, stent graft migration, stent graft fracture, stent graft kinking, insufficient stent graft expansion and stent graft embolism.Delivery system specific events that could be associated with clinical complications include bond joint failures, detachment of parts, incompatibility with accessory devices, premature deployment, inaccurate deployment, failure to deploy, high deployment forces, delivery system kinking, no visibility under fluoroscopy, inability to track to target location, and blood leakage from delivery system (hemostasis).Stent graft size selection: special care must be taken to ensure that the appropriate size flair endovascular stent graft is selected.The stent graft body diameter should be approximately 1 mm larger than the synthetic av access graft diameter.
|
|
It was reported that during deployment of the stent the health care provider allegedly experience resistance during treatment in the brachial artery graft via the basilica artery.It was further reported that the catheter was removed without incident and was identified to be broken.Reportedly, a pta balloon was then used to continue the procedure, and angioplasty was performed.There was no reported patient injury.
|