• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTURESNARE, 90 DEGREE STRAIGHT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTURESNARE, 90 DEGREE STRAIGHT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number AR-6060-90
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 04/24/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainants event is typically caused by use of excessive force and/or prying/leveraging with the tip fully exposed on the device during use.This is the (b)(4) complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Device requested but not yet received.
 
Event or Problem Description
It was reported that during a rotator cuff repair procedure, the surgeon was attempting to use a 90° suture snare.He caught a suture with the snare and as he pulled, he felt resistance.He continued to pull and the tip of the wire broke off into the patient's joint.The surgeon attempted to retrieve the tip but was unsuccessful as he was unable to locate the tip.The fragment remained in patient body.The surgeon took a post-op x-ray however, the tip could not be seen.
 
Additional Manufacturer Narrative
Patient demographics (age at time of event, date of birth, gender, weight) were requested but not provided.No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.This is a follow-up submission due to device evaluation.Device history record review revealed nothing relevant to this event.Complaint confirmed.The device met all material specification as received.The evaluation of the returned device revealed that the nitinol wire at the distal tip is broken off.Complainants event is typically caused by use of excessive force and/or prying/leveraging with the tip fully exposed on the device during use.This is the first complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a rotator cuff repair procedure, the surgeon was attempting to use a 90° suture snare.He caught a suture with the snare and as he pulled, he felt resistance.He continued to pull and the tip of the wire broke off into the patient's joint.The surgeon attempted to retrieve the tip but was unsuccessful as he was unable to locate the tip.The fragment remained in patient body.The surgeon took a post-op x-ray however, the tip could not be seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURESNARE, 90 DEGREE STRAIGHT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key6571648
Report Number1220246-2017-00172
Device Sequence Number11015498
Product Code LXH
UDI-Device Identifier00888867204638
UDI-Public00888867204638
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date (Section B) 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/01/2021
Device Catalogue NumberAR-6060-90
Device Lot Number10072434
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/25/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/17/2017
Supplement Report FDA Received Date06/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-