• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLAIR ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLAIR ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FAF09050
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an attempt to deploy the endovascular stent graft in an upper arm, the outer catheter sheath and the handle of the delivery system allegedly broke.Reportedly, another endovascular delivery system was used and the procedure was completed.There was no reported patient injury.
 
Additional Manufacturer Narrative
The device history records have been reviewed with special attention to the subassemblies, material review reports, raw material testing, manufacturing process, and quality control inspection.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint that has been reported for this corporate lot number to date.Visual/dimensional evaluation: the device was returned in two separate segments.The stent graft was deployed and not returned.The touhy was tight and the red safety handle was not returned.The outer catheter had a complete circumferential break 5.8cm from strain relief.No anomalies were noted to the outer catheter.Functional/performance evaluation: functional testing could not be performed, as the outer catheter was broken.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the investigation was confirmed for break as the outer catheter had a complete circumferential break.Per the reported event details, it was noted that the delivery system was advanced without a sheath.There is a specific precaution listed in the ifu that indicates that the safety and effectiveness of this device have not been established when used without a sheath.Therefore, it is possible that procedural issues contributed to the reported event.However, the definitive root cause could not be determined.Labeling review: the current ifu (instructions for use) state: indication for use: the flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Warnings: the stent graft (implant) cannot be repositioned after total or partial deployment.Once partially or fully deployed, the flair endovascular stent graft cannot be retracted or remounted onto the delivery system.Device removal after deployment can only be done with a surgical approach.Precautions: the safety and effectiveness of this device have not been established when used around a tight bend in a looped av graft.It is recommended to access the av graft at the venous side of the av graft or at the apex.Careful attention should be paid to ensure the device is appropriately sized to the actual graft diameter, taking into account any change to the stated graft diameter that may have resulted from previous interventions.The appropriate length device(s) should be selected so that the stent graft extends beyond the stenosis into at least 10 mm of the non-diseased graft towards the arterial inflow and into the non-diseased vein approximately 10 mm beyond the stenosis.During deployment of the stent graft, the entire length of the delivery system should be kept as straight as possible in order to mitigate the potential for distal stent graft misplacement.After full stent graft deployment, wait a few seconds to allow for complete device expansion before removing the delivery system over the guidewire.Stent graft dislodgement has been reported during removal of the delivery system; therefore, careful attention should be paid during this portion of the procedure to prevent such occurrences.In case of placement of two stent grafts (overlap), use the same device diameter in both cases.If a flared device is used to overlap, do not deploy the flared end inside the first stent graft.Ensure a minimum 10 mm overlap of the devices.Potential complications and adverse events: stent graft specific events that could be associated with clinical complications include stent graft misplacement, stent graft migration, stent graft fracture, stent graft kinking, insufficient stent graft expansion and stent graft embolism.Delivery system specific events that could be associated with clinical complications include bond joint failures, detachment of parts, incompatibility with accessory devices, premature deployment, inaccurate deployment, failure to deploy, high deployment forces, delivery system kinking, no visibility under fluoroscopy, inability to track to target location, and blood leakage from delivery system (hemostasis).Stent graft size selection: special care must be taken to ensure that the appropriate size flair endovascular stent graft is selected.The stent graft body diameter should be approximately 1 mm larger than the synthetic av access graft diameter.
 
Event or Problem Description
It was reported that during an attempt to deploy the endovascular stent graft in an upper arm, the outer catheter sheath and the handle of the delivery system allegedly broke.Reportedly, another endovascular delivery system was used and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLAIR ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6572862
Report Number2020394-2017-00468
Device Sequence Number92184
Product Code MIH
UDI-Device Identifier04049519008875
UDI-Public(01)04049519008875(17)191110(10)ANAY1421
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P060002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2019
Device Model NumberFAF09050
Device Catalogue NumberFAF09050
Device Lot NumberANAY1421
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2017
Initial Date Received by Manufacturer 04/20/2017
Supplement Date Received by Manufacturer06/26/2017
Initial Report FDA Received Date05/17/2017
Supplement Report FDA Received Date07/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-