| Model Number |
MICL13.7 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Pain (1994); Blurred Vision (2137); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Another similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter stated the surgeon implanted a 13.7mm micl13.7 implantable collamer lens, -10.0 diopter; in the patients left eye (os) on (b)(6) 2017.The reporter stated the lens was explanted on (b)(6) 2017 due to excessive vaulting; elevated intraocular pressure; narrowing of the angle; angle closure and shallowing of the anterior chamber.The patient experienced blurry vision and pain.The lens was not exchanged for another lens.The reporter stated the cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the lens was returned in a specimen cup.Visual inspection of the returned product found a piece of lens optic and haptic torn off and missing.A piece of the other haptic was torn off and missing.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The reporter indicated a shorter lens (13.2mm) was implanted on (b)(6) 2017 and the lens exchange resolved the problem.The lens vault was good and the intraocular pressure was normalized.The patient is doing good, with a visual acuity of 20/20.(b)(4).
|
| |
|