| Model Number |
37592-65 |
| Medical Device Problem Code |
Missing Value Reason (3192)
|
| Health Effect - Clinical Code |
Missing Value Reason (3192)
|
| Date of Event |
04/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We received the feedback from our distributor, that the label of two screws with the art.No 37592-65 and the lot no.18/772125 differs from the content.The investigation showed that five labels got reprinted on 3rd of november 2016.After checking the data in the erp system, three facilities resp.Distributors have received these implants.There is no risk for any patient or surgeon or third party.The non-conformance gets identified at refilling the non-sterile screws in the screw caddy, which happens prior to the cleaning and disinfection process and so prior to the use in the o.R.
|
| |
|
Event or Problem Description
|
|
We received the feedback from our distributor, that the data on the label differs from the content of the non-sterile implant packaging.The information on the label showed a cancellous locking screw 5.9mm and the content was a stabilization locking screw 2.4mm.The issue was identified at goods-in inspection from our distributor, so there was no risk associated with a patient, surgeon or third party.The screws get delivered non-sterile to the facility, so the non-conformance gets identified latest at refilling the sterilization tray prior to the cleaning and disinfection process and prior to any surgery.
|
| |
|