| Model Number |
E SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical (b)(6); the reported malfunction was observed and attributed to a loose system interconnect flex cable.The system interconnect flex cable was replaced to remedy the problem.The device was recertified and returned to the customer analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|