• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FOUNDATION KNEE; FEMUR,POSTERIOR STABILIZED, OPEN BOX, SIZE6, LT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FOUNDATION KNEE; FEMUR,POSTERIOR STABILIZED, OPEN BOX, SIZE6, LT Back to Search Results
Catalog Number 213-01-006
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/19/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to instability; loose components.
 
Additional Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was instability.The in-vivo length of patient service for the implant was 11.8 months.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was the patient had instability issues from loose components.The scope of this investigation is limited without having the explanted parts available to djo surgical for evaluation.Other conditions relating to this event could not be determined with confidence.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION KNEE
Common Device Name
FEMUR,POSTERIOR STABILIZED, OPEN BOX, SIZE6, LT
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6576965
Report Number1644408-2017-00388
Device Sequence Number16276917
Product Code JWH
UDI-Device Identifier00888912110266
UDI-Public(01)00888912110266
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K964008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number213-01-006
Device Lot Number837A1060
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/19/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/18/2017
Supplement Report FDA Received Date06/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
360-11-506, LOT 59604140
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient Weight86
-
-