• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/19/2017
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose.Once the evalutaion will be performed, a follow-up medwatch report will be submitted.
 
Event Description
It was reported that at the beginning of the surgery, while locking the robot into position using the hydraulic stabilization system, it was observed that one of the three hydraulic feet hit the ground before the two others.Upon further inspection after the case, it was observed that the subject hydraulic foot was bent.
 
Manufacturer Narrative
Hydraulic stabilization system has been inspected for investigation purpose.The causes identified are user error and unclear ifu.The ifu states that the feet should be raised to detach the patient but only in the emergency procedure paragraph.However, it does not specifically warn the user to raise the feet during transport nor does it say how high to raise the feet.Corrected data: date received by manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key6577682
MDR Text Key75520287
Report Number3009185973-2017-00612
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 11/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROSA 2.5.8
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/19/2017
Initial Date FDA Received05/19/2017
Supplement Dates Manufacturer Received11/07/2017
Supplement Dates FDA Received11/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/14/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-