Catalog Number PNML6F088904 |
Device Problems
Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311); Split (2537)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the neuron max 6f 088 long sheath (neuron max) was fractured approximately 3.0 cm from the hub and kinked approximately 47.0 cm from the hub.Conclusions: evaluation of the returned device confirmed that it was fractured at the distal end of the strain relief.This damage may have occurred due to forceful handling of the neuron max during removal from the packaging.Further evaluation revealed the neuron max was kinked.This damage was likely incidental and likely occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
While prepping a neuron max 6f 088 long sheath (neuron max) for the procedure, the physician noticed that the neuron max was split at the strain relief.The damaged neuron max was found prior to use and therefore, the neuron max was not used in the procedure.The procedure was completed using another neuron max.
|
|
Search Alerts/Recalls
|