Analysis was approved for the returned generator.When received, the programming data was downloaded from the generator and reviewed.High impedance was first observed 4 months after implant and occurred intermittently through the remainder of the implant time of the device.Visual examination of the generator did not identify any abnormalities with the external part of the generator.The generator was interrogated, and system diagnostic tests were performed and returned results within the normal limits.Electrical evaluation showed that the device performed according to functional specifications.There were no performance or any other type of adverse condition found with the pulse generator.Analysis was approved for the returned lead.The lead body was returned in one portion.Note that the electrodes were not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.Set screw marks were observed towards the end of the connector pin, indicating the connector pin had not been fully inserted into the generator.No other sets of set screw marks were observed.Canted spring indentations were also observed on the rear end of the small front o-ring, indicating insufficient mechanical contact between the generator and connector ring.The position of the set screw marks and canted spring indentations indicate that there may have been a suspect electrical connection to the lead.No discontinuities were identified during the continuity check of the lead.There was no evidence to suggest an anomaly with the returned portion of the lead.Incomplete insertion of the connector pin may have been a potential cause for the observed high impedance condition.No additional relevant information has been provided to date.
|