• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 05/01/2017
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass procedure, during prime, the right angle luer in the one way line manifold cracked and leaked.*no patient involvement, *product was changed out, *surgery completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on may 21, 2017.(b)(4).Evaluation of the returned sample found that the l connector was cracked along the part.Replication testing found that this crack appears on the part when the l connector is over tightened on a port.During setup of the circuit, the l connector had likely been over-tightened, causing it to crack.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
125 blue ball road
elkton MD 21921
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key6580806
MDR Text Key75603928
Report Number1124841-2017-00088
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberVD06
Other Device ID Number(01)00699753450820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-