| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause of malfunction: user had an inaccurate reference.(b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with test results from results obtained of 68 mg/dl.The customer's expected fasting blood glucose test result range is 80 - 110 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was performed by the customer fasting and produced test results of 103 mg/dl using truetrack meter and 115 mg/dl using truetrack meter.The product is stored according to specification.The test strip lot manufacturer's expiration date is 12/31/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|