| Model Number |
48233540 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the surgeon was undertaking a spinal procedure and after everything had been locked down the surgeon felt that the construct was loose.He felt around to understand what was happening and as he did so the tulip head came off the screw in the surgeon's hand.The surgeon was able to remove the shaft of the screw and insert a new one.The case was completed successfully and there were no adverse consequences for the patient.There was a couple of minutes delay while the new screw was inserted.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, device history review, complaint history review, risk assessment.Result: the customer reported was confirmed via visual inspection.The tulip head was disengaged from the main thread body.No relevant manufacturing issues were identified as all units met stryker specifications.Conclusion: due to the multifactorial nature of the event, a definite root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was undertaking a spinal procedure and after everything had been locked down the surgeon felt that the construct was loose.He felt around to understand what was happening and as he did so the tulip head came off the screw in the surgeon's hand.The surgeon was able to remove the shaft of the screw and insert a new one.The case was completed successfully and there were no adverse consequences for the patient.There was a couple of minutes delay while the new screw was inserted.
|
| |
|
Search Alerts/Recalls
|