MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MELODY TRANSCATHER PULMONARY; PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MELODY TRANSCATHER PULMONARY; PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number PB1018
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Rupture (2208); Pericardial Effusion (3271)
Event Date 04/18/2017
Event Type  Injury  
Manufacturer Narrative
Product analysis: the product remains implanted, therefore no product analysis can be performed.Conclusion: without return of the product, no definitive conclusions could be drawn regarding the clinical observation.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that following the implant of this transcatheter pulmonary bioprosthetic valve, a balloon aortic valvuloplasty (bav) was performed that caused a rupture distal to the valve.Flow into the pericardium was then noted.A 39 mm covered stent was deployed and successfully contained the rupture.A second valve was implanted valve-in-valve to provide a competent pulmonary valve.A pericardial drain was placed and resulted in 170 ml of blood drained.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MELODY TRANSCATHER PULMONARY
Type of Device
PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key6582815
MDR Text Key75690556
Report Number2025587-2017-00834
Device Sequence Number1
Product Code NPV
UDI-Device Identifier00643169588196
UDI-Public00643169588196
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/23/2018
Device Model NumberPB1018
Device Catalogue NumberPB1018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2017
Initial Date FDA Received05/22/2017
Supplement Dates Manufacturer Received04/28/2017
Supplement Dates FDA Received10/04/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age29 YR
Patient Weight100
-
-