• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. MEDEX¿ SMALL BORE STOPCOCKS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. MEDEX¿ SMALL BORE STOPCOCKS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number MX5341LN
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Seizures (2063); Underdose (2542)
Date of Event 04/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is currently being evaluated; smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.
 
Event or Problem Description
It was reported that a medex¿ small bore stopcock was in use with a patient when it was observed that there was medication leaking.The patient was assessed at 8:00 am and the nurse observed that there was a large amount of fluid under the patient's right leg on the sheets.The patient was being infused with propofol through the patient's right femoral central venous catheter (cvc).The cvc line was attached to a stopcock, which also managed continuous fentanyl and levophed combination delivery.Attachments were inspected and were found to be secure and without leaks.It was observed that the back of the stopcock bundle had propofol residue on it and was confirmed with a chlorhexidine gluconate wipe to clean the area.After the area was cleaned, it was observed that white fluid appeared in large drops.The patient had seizure activity at 6:00 am on the next day due to the incident and propofol infusion dosage was increased to treat the seizure activity.No permanent injury was reported.
 
Additional Manufacturer Narrative
Four medex¿ small bore stopcocks were returned for investigation.Visual inspection revealed obvious damage/cracking of the returned device.The observed damage/cracking was consistent with the device being exposed to chemicals for a prolonged period of time.The design of the device is not intended to be impervious to all drugs and combinations of drugs.The observed damage/cracking would not have occurred during the production of the device.Additionally, the manufacturing facility performed a quality inspection of (b)(4) samples and functionally tested an addition (b)(4) devices and no issues were noted during inspection.Investigation determined that the root cause of the reported event was due to the improper use of the device.
 
Additional Manufacturer Narrative
One stopcock was returned for evaluation.Visual inspection discovered there was obvious damage/cracking that occurred.The damage observed was consistent with prolonged chemical exposure and would not occur during normal production of the device.Based on the evidence, the root cause was attributed to the use of the product in a manner inconsistent with its instructions for use.There was no evidence to indicate that there was a defect due to the manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDEX¿ SMALL BORE STOPCOCKS
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key6582892
Report Number3012307300-2017-01089
Device Sequence Number13477353
Product Code FMG
UDI-Device Identifier10351688506771
UDI-Public10351688506771
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date10/27/2019
Device Catalogue NumberMX5341LN
Device Lot Number3312279
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 05/03/2017
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/22/2017
Supplement Report FDA Received Date05/31/2017
06/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-