Catalog Number DR8056
Medical Device Problem Codes Break (1069); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/30/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Visual/microscopic inspection: the balloon size for this product is printed on the balloon hub of the catheter and identified the returned sample as a 5 mm x 6 cm balloon.The device was examined under magnification (25x), and a partial circumferential break was noted at the proximal balloon butt glue joint.The balloon appeared to have been previously inflated and no fiber disturbance was noted.No other anomalies were observed to the device at this time.Functional/performance evaluation: no functional testing could be performed due to the break at the proximal balloon glue joint.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the device was returned.The device was examined under magnification and a partial break was noted to the glue joint between the outer catheter and the balloon, confirming the investigation for a break.The investigation is inconclusive for the reported rupture as functional testing could not be performed, due to the break.The definitive root cause for the reported rupture or identified break could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high-quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip breakage or balloon separation.Do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/introducer sheath as a single unit.Use: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place and inflate the balloon to the appropriate pressure.Apply negative pressure to fully evacuate fluid from the balloon.Confirm that the balloon is fully deflated under fluoroscopy and that no contrast is left in the balloon.(b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a calcified lesion in the left common iliac artery the pta balloon allegedly ruptured at 6 atm on the second inflation attempt.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA BALLOON DILATATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key6583518
Report Number2020394-2017-00490
Device Sequence Number13639731
Product Code DQY
UDI-Device Identifier00801741058523
UDI-Public(01)00801741058523(17)200308(10)93QB0100
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/08/2020
Device Catalogue NumberDR8056
Device Lot Number93QB0100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2017
Initial Report FDA Received Date05/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight81
-
-