| Model Number |
6955252 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/27/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for investigation.A review of the radiograph provided confirmed screw and spacer migration.The root cause of this event has not been determined.Labeling review: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation" still in-situ.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017 a (b)(6) y/o female underwent a vertebral body replacement.On (b)(6) 2017 follow up radiographs showed the interfixated screw backing out and cage migration at l4-l5.As per initial reporter no revision procedure is planned at this time.
|
| |
|
Search Alerts/Recalls
|