• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G34505
Device Problem Positioning Failure (1158)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation is still in progress.
 
Event Description
Description of event according to initial reporter: the legs of the filter would not deploy.They were stuck together.They resheathed it and removed it from the patient.Patient outcome: no adverse effects on the patient due to this occurrence.
 
Manufacturer Narrative
Exemption number e2016032.(b)(4).Name and address for importer site: (b)(4).Summary of investigational findings: a jugular introducer and a celect-pt filter were returned.Investigation found the filter manufactured according to specifications and revealed no damages.Based on these findings and since no imaging was provided, the exact reason, why the legs of the filter would not deploy cannot be determined.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended.Cook medical will continue to monitor for similar events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT® PLATINUM NAVALIGN UNISET VENA CAVA FILTER SET
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov DK-46-32
DA   DK-4632
56868686
MDR Report Key6583835
MDR Text Key75718673
Report Number3002808486-2017-01108
Device Sequence Number1
Product Code DTK
UDI-Device Identifier10827002345055
UDI-Public(01)10827002345055(17)190913(10)E3493641
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG34505
Device Catalogue NumberIGTCFS-65-1-UNI-CELECT-PT
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Distributor Facility Aware Date05/01/2017
Device Age7 MO
Date Manufacturer Received05/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
-
-