• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Codes No Patient Involvement (2645); No Information (3190)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, a 980 ventilator generated a serial number mismatch error message.Although requested, it is unknown if the p atient was connected to the ventilator at the time of the reported event.The service engineer (se) inspected the device and verified the reported issue.The se replaced the universal serial bus (usb).The se performed calibrations and extended self-testing on the device and all tests passed.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The reported issue occurred during service.The ventilator was not in use on a patient at the time of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key6584069
Report Number8020893-2017-06197
Device Sequence Number16309105
Product Code CBK
UDI-Device Identifier10884521201385
UDI-Public10884521201385
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K131252
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980
Device Catalogue Number980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2017
Initial Date Received by Manufacturer 04/24/2017
Supplement Date Received by Manufacturer10/02/2017
Initial Report FDA Received Date05/23/2017
Supplement Report FDA Received Date10/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-