| Model Number |
980 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, a 980 ventilator generated a serial number mismatch error message.Although requested, it is unknown if the p atient was connected to the ventilator at the time of the reported event.The service engineer (se) inspected the device and verified the reported issue.The se replaced the universal serial bus (usb).The se performed calibrations and extended self-testing on the device and all tests passed.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The reported issue occurred during service.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|